Telecordia GR-49-CORE for outdoor network interface devices
Don't be fooled. Test 5.3 is harder than it sounds because it requires humidity to be controlled below 10°C, which is very hard on a chamber. Also, for test 5.3, we don't recommend our benchtop series due to the requirement for humidity to be controlled during temperature transition, starting at 4°C.